Pensky-Martens Closed Cup Flash Point Tester LB-13PFP

Pensky-Martens Closed Cup Flash Point Tester LB-13PFP

Pensky-Martens Closed Cup Flash Point Tester LB-13PFP accurately determines the flash point of various substances. It adopts manual ignition method. Utilizing a mercury-in-glass thermometer, the tester offers multiple scales for comprehensive flash point determination. With a compact and exquisite design, it boasts a stainless-steel table board, combining reliability with sophistication.

Get Quote Download Catalog

Specifications

Ignition method Manual
Ambient Temperature 15 to 35 °C
Relative Humidity ≤ 85%
Compliance ASTM D93
Flash Point Determination Mercury-in-glass thermometer
Scale -5 °C to 110 °C, division 0.5 °C
Scale 20 °C to 150 °C, division 1 °C
Scale 90 °C to 370 °C, division 2 °C
Oil Cup Inner diameter: 50.7 mm to 50.8 mm
Depth: 55.7 mm to 56.0 mm
Scribed line depth of capacity: 33.9 mm to 34.3 mm
Capacity of testing oil: about 70 ml
Stirring Rate Procedure A: 90 to 120 rpm, Procedure B: 250 ± 10 rpm
Igniting source Gas
Heating Device Continuously adjustable from 0 W to 600 W
Heating Rate 1 to 12 °C/min (Controllable and adjustable)
Power Supply AC 220 V ± 10%, 50 Hz
Power ≤ 650 W
Dimension (W × D × H) 370 × 320 × 300 mm
Packaging Dimension (W × D × H) 650 × 450 × 360 mm
Gross Weight 19 Kg

Features

  • Adjustable heating power
  • Accurate Oil Cup Design for Pensky-Martens Closed Cup Flash Point Tester
  • Variable stirring rates
  • Controllable heating rate
  • Efficient and user-friendly
  • Sturdy construction

Applications

Pensky-Martens Closed Cup Flash Point Tester is used to determine flash point of petroleum products in research, chemical, petroleum, oil and other industries for classification, shipping and safety regulation.

Get Quote

Pensky-Martens Closed Cup Flash Point Tester

Ignition method Remote control
Ignition gas Pipe gas or LPG
Pressure ≤ 10 kPa
Measuring range 40 °C ~ 300 °C
Flame Pilot flame
Ignition method Automatic
Flame Pilot flame
Temperature sensor Glass pt100
Compliance ASTM D93
Ignition method Automatic
Measurement principle Thermocouple differential detection
Temperature Measurement Range Room temperature to 350 ℃
Resolution 0.1 ℃
Temperature Range Room Temperature to 350℃
Stirring Speed Grade A:125 r/min, Detection Temp Greater than 110℃
Temperature Control Accuracy ±0.1℃
Environmental Change Range 50℃ to 450℃
×

Ask Quote